Public Lab Research note


Hydro-powered DIY FTIR Spectrometer

by bryanchua | December 18, 2018 18:44 18 Dec 18:44 | #17968 | #17968

Hi,

My friends and I came up with this idea in a makerspace class. Does anyone know if such an idea would work and what materials would be needed? The end goal is to make something that could detect for chemical pollutants like mercury and microbiological contaminants like E. coli. It would be deployed in rural areas with little or no WiFi or cellular networks and the LoRaWAN chip will transmit water quality data reliably and anonymously so people on the ground would not be persecuted by illegal gold mining or drug syndicates who pollute their water. The hydro-powered motor powers the floating device and recharges the battery to power the Fourier-Transform Infrared (FTIR) spectroscopy module that will periodically measure the water.


7 Comments

I like the idea. But why are you selecting ftir?

Is this a question? Click here to post it to the Questions page.

Reply to this comment...


I've never seen FTIR used for mercury detrmination. Usually, that is a variation of AA.

The instrument set up you show is for transmission. Probably, reflectance would be needed instead. Either MIR or ATR. That leads to a different situation. Then, how do you ID a specific microbe by FTIR? The common instruments can't do that. Is there a reference you have? Or a different technique?

Please keep investigating and good luck.

Is this a question? Click here to post it to the Questions page.

Reply to this comment...


Thanks for the feedback @Ag8n !

We used FTIR because we saw that people have made DIY FTIR spectrometers and we also read these articles where FTIR was used for mercury detection and E. coli detection respectively:

https://www.ncbi.nlm.nih.gov/pubmed/22560281 https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3458592/

Reply to this comment...


Just realize, these methods are far from being implemented. The mercury one is " proof of concept", which means it still needs troubleshot, validated, etc. to be a standard method. It also uses a mix of methods. SPE( solid phase extraction ) either takes a robot or a person to do. The work and can be touchy, depending on the test done. Usually, the sample is spiked and recovery calculated.

The e. Coli method is based on a detector that is liquid nitrogen cooled. Not your run of the mill instrument. They tend to be much more expensive than a homemade type ftir.

Keep up the research!

Reply to this comment...


hello, I started a research with sensing of microplastics in ocean waters. [1]

As the idea is for the sensor to be embedded in the float, we don't want to use reagents in the sample sockets, so we discarded UV and started researching FTIR and polymer detection spectrum, a lot with the ATR technique, but we thought about doing just transmittance in the analysis cuvette.

here an explanatory video of the project with subtitles https://www.youtube.com/watch?v=ienOsKmb7cA

[1] tortugaguardian.org / https://publiclab.org/wiki/tortuga-guardian

Is this a question? Click here to post it to the Questions page.

Thank you for sharing this gorgeous video with subtitles! I love this project. There's going to be a OpenHour on environmental defense in July in Spanish -- perhaps you could join and share? https://publiclab.org/openhour#Monday,+July+6/lunes,+6+de+julio:+Public+Lab+Hora+Abierta CC @xose @imvec

Is this a question? Click here to post it to the Questions page.


Reply to this comment...


An impressive step forward.

I couldn't tell, from the visual, if the instrument used FTIR transmittance or FTIR reflectance.
Could you please clarify? Also, does the instrument use FTIR or NIR?

By the way, the instrument cost is very good.

Thank you.

Is this a question? Click here to post it to the Questions page.

Reply to this comment...