Public Lab Research note


The user 'dhaffnersr' has been placed in moderation and will not be able to respond to comments.
  • 0

Empirical Verification Of Slit Width For The DH4.0 v4 Spectrometer Section II

by dhaffnersr |

Verification through quality control check procedures

Using my Premeire Professional binocular microscope, I verified my new slit template and design. The razor blades are now mounted on a piece of polycarbonate for strength and rigidity, which upon microscopic inspection reveals that indeed, both edges appear horizontal in respect to their positions.

Exact measurements are noted on the title image. (note_all original data is available upon request.)

Slit width length according to specifications are; 1) Length = 34mm, Width = 0.06mm 2) Template dimensions: Length = 63.49mm, Height = 44.44mm 3) Template thickness: 1.69mm

Razor blades used: Gillette utility knife blades (standard size)

laser_scan_5_sept_9_FIG_1.png

laser_scan_5_sept_9_FIG_2.png

I measured my spectral bandwidth using the FWHM value of my 150mW laser data as shown in figure.2, the reason for this is because my laser's performance at specifications are highly reliable and very repeatable, so I trust the data. I understand that there are a few here at Plab, that question my results and methods, what I have to say to those individuals is, prove my data incorrect, don't just make innuendos, get your spectrometer out and show me how its done, talk is cheap.

References

http://publiclab.org/notes/dhaffnersr/09-09-2016/empirical-verification-of-0-05mm-slit-width-of-dh4-0-v4-spectrometer



spectrometer spectrum-matching spectrography image-compositing image-processing citizen-science field-test spectralworkbench grassroots-science weird-science plots ultraviolet database compact-fluorescent testing spectral-analysis science refinement google google-maps google-earth tests led chlorophyll metadata data-collection data popular-science museum-of-science leds quantitative datalogger civic-science slit-width specification research-notes electrochemical wavelength technique purelab reference hopewell city-point city-point-hopewell-va dhaffnersr david-h-haffner-sr spekwin32 david-h-haffner haffner david ethanol quartz graphs gnu-plot 400nm 400-nm-led technical technical-specs technical-specifications testing-uv-led images-led-testing images-led-testing-400-nm-uv images-test-led-400nm-ultra-violet uv-spectrum-200-400-nm-leds-testing-test-results quartz-cuvettes quartz-cuvette 400nm-led-thru-quartz-cuvette-test electrical-engineering google-search meta-tag meta-tags chlorophyll-testing cool-stuff very-cool-stuff relevant uv-vis uv-vis-ethanol b aspirin uv-vis-chlorophyll dye dyed eosin eosin-y extract portable-laser emission-wavelength wavelengths generic-laser uv-vis-sodium-hydroxide redox width-of-dvd-grating slit-length distance-of-slit bandwidth methylene-blue blank-distilled-water aries emission-wavelength-for-rhodamine-b image-raman-spectrometer google-image-rhodamine-b-samples google-image-raman-spectrometer-homemade plots-spectrometry optical-spectroscopy circular-dichroism-spectroscopy fluorescence-dye photophysik led-testing led-test .