Where can I report this?
The blue bar appears to need to be negative, and the calibration is therefore not stored.https://spectralworkbench.org/spectrums/136374
Is there a way of adding a calibration with user-defined calibration positions?
404.7 - H-line - violet
435.8 - G-line - blue
546.1 - green
578.2 - yellow-orange
I found there's a manual calibration URL I could use:
https://spectralworkbench.org/spectrums/calibrate/136374?x1=165&w1=546&x2=189&w2=579
This fixed the issue for me.
(The X position of the bars is simply found by hovering over the line chart when it's uncalibrated - which gives the X value from the leftmost end of the spectral image.)